Fitfunktion Temperature-Kategorie fittet nicht #100
Labels
No Label
Kind/Breaking
Kind/Bug
Kind/Crash
Kind/Documentation
Kind/Enhancement
Kind/Feature
Priority
Critical
Priority
High
Priority
Low
Priority
Medium
Priority
Very low
Reviewed
Duplicate
Reviewed
Invalid
Reviewed/Won't Fix
Status
Need More Info
Status
On Hold
Status
Stale
Type/BDS
Type/DSC
Type/Fit
Type/General
Type/NMR
No Milestone
No project
No Assignees
2 Participants
Notifications
Due Date
No due date set.
Dependencies
No dependencies set.
Reference: IPKM/nmreval#100
Loading…
Reference in New Issue
Block a user
No description provided.
Delete Branch "%!s()"
Deleting a branch is permanent. Although the deleted branch may continue to exist for a short time before it actually gets removed, it CANNOT be undone in most cases. Continue?
Die Vorschau für die Startwerte beim VFT sieht okay aus, der Fit ändert aber nur die Fehler und passt die Parameter selbst nicht an. ErrorLog behauptet, es würde alles funktionieren. Auch nachdem ich es aus und wieder angeschaltet habe. Arrhenius klappt auch nicht, andere Funktionen (Basic/Straight Line ausprobiert) funktionieren gut.
Der Fit funktioniert, die Daten sind das Problem. Weil du so viele Punkte bei 298 K und den anderen höheren Temperaturen hast, ist der Bereich überproportional wichtig. Der Fit hört auf, sobald er diesen Bereich gut genug anpassen kann, was ziemlich unmittelbar passiert.
Kann ich manuell das Konvergenzkriterium für den Fit verschärfen?
Nein, das geht nicht. Hast du schon versucht, die Gewichtungen der Punkte zu ändern?
Für dieses datenset noch nicht.
Nicht direkt zumindest. Ich habe dann die Daten mit Evaluate Expression zwischen y=np.log(y) und y=exp(y) hin und her geschubst, bis ich eine Gerade fitten konnte und die dann wieder zurücktransformiert.
Hoffentlich mit Transformation von T zu 1/T, ansonsten ist die Gerade ganz großer Unsinn.