Markus Rosenstihl 5b52fb3e5a Spectrometers with new computers show increasingly timeouts.
This can  be because the timer in the backen is not synchronized with the pulseblaster.
When the timer starts running >0.5s before the pulse program is started, a timeout could occur.
The current value of additional 0.5s is changed to 2s. This timer has nothing to do with the trigger line.

- fixes T21